APC-1106T High-precision Laser Dust Particle Counter
关键词:
所属分类:
咨询热线:
APC-1106T High-precision Laser Dust Particle Counter
APC-1106T High-precision Laser Dust Particle Counter
Product Overview
Light source: semiconductor laser
Sampling flow rate: 28.3L/min
Built in flow sensor, automatic closed-loop flow control
Particle size channels: 0.1, 0.2, 0.3, 0.5, 1.0, 5.0 μ m
Display: Large screen LCD touch screen (LCD)
Power supply: Lithium battery rechargeable 16.8V
Adopting DC brushless fan and PWM regulation control
Data storage: capable of storing at least 20000 sets of data
Printing function: Thermal printer
Product Features
This instrument is a successful product developed in collaboration with relevant universities and research institutions in China based on the international standard ISO 14644-1:2015. The overall body of the product is made of 316L stainless steel shell. The laser sensor adopts a full semiconductor laser and a semiconductor photodiode receiver to ensure the stability of the light source and the accuracy of the signal receiver. Greatly reducing the stray light inside the scattering cavity and improving the signal-to-noise ratio of the sensor. Adopting LCD touch screen control, the display data is clear and the operation is convenient. The sampling air pump is imported from the United States and is equipped with a pressure differential sensor imported from the United States for closed-loop control of the flow rate, ensuring that the flow rate remains stable within the set range. This instrument can automatically determine various purification levels based on the standards selected by the customer.
Scope of application
This product is widely used in medical, pharmaceutical, electronic, chemical, optical, food, cosmetics, biological products, aerospace, and various testing institutions for the detection of dust particles
Main technical parameters:
light source
Semiconductor laser, with a lifespan of over 10 years>30000 hours
sampling flow
28.3L/min (1ft3) ± 5% can maintain a constant flow rate according to the length of the sampling tube resistance
display mode
155 × 85mm high-resolution LCD touch screen display (LCD) Chinese interface, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter), three-level permission management, audit tracking function.
Power supply/working power supply
Maximum power consumption: 20W. Built in long-lasting lithium battery: capable of continuous measurement and operation for 6 hours. External charger can be quickly charged (<4 hours). The charger provides power from 100V to 220V, 50/60Hz
working hours
6 hours
Counting Mode
Cumulative value, difference value, concentration value
test method
Single, repetitive, continuous, computational, remote
unit conversion
The unit can be converted into m3
Particle size channel
Simultaneously count six particle sizes of 0.1, 0.2, 0.3, 0.5, 1.0, and 5.0 μ m Particle size can be customized according to customer needs
sampling period
1~99999 (S) Delay count: 0~99 (S) Self purification time: ≤ 10 (min)
working environment
Temperature: 5-35 ℃ Relative temperature: 20-90% RH, no condensation Atmospheric pressure: 86-106KPa
Temperature and humidity:
Choose (1) Temperature: -15~50 ℃± 1 ℃ (2) Humidity: 0-98% RH ± 5%
UCL settings
Sampling points (A): Set from 0 to 99 points. Sampling frequency per point: (L) 0-99 times Set measurement position: 0-99999999
UCL report
Compliant with ISO14644-1, GMP (2010), FS 209E
data storage
It can store 20000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, data location port), and the data will not be lost after power failure.
communication interface
RS232, 9600 baud rate, USB interface
Alarm Settings
The instrument is equipped with a level alarm function, which can sound an alarm when the cleanroom exceeds the standards of 100, 1000, 10000, 10000, 30000, 1000000. At the same time, it can switch between four measurement standards, A/B/C/D, static and dynamic, in accordance with the new version of GMP standards
Zero Count
Complies with JJF-1190-2008 Dust Particle Counter Calibration Specification and ISO/FDIS21501-4:2018, with less than one count every five minutes
Overlap error
At 5%, 2000000 particles per cubic foot
Print function
thermal printer
weight
7.0kg
calibration
It can meet the JJF-1190-2008 dust particle counter calibration specification, GB/T6167-2007 dust particle counter performance test method, and ISO/FDIS21501-4:2018 (E) clean room light scattering dust particle counter calibration method. It can be traced back to Shanghai Institute of Metrology and Measurement Technology, self calibration or third-party national metrology institutions calibration, and can also be traced back to the National Institute of Standards and Technology (NIST) in the United States. Our company has passed the national metrological standard building assessment, and our standard dust particle counter has also participated in the national dust particle counter value comparison with the participation of various provincial metrology institutions. It is used as a standard precision particle counter for relevant metrology and testing institutions to carry out dust particle counter calibration work.
Standard Configuration
Professional power sampling head, sampling bracket with sampling tube, built-in printer, aluminum alloy box, operation manual, factory inspection report, zero count filter, charger, sampling rack, printing paper
Overall dimensions
230 * 280 * 230 (mm) (excluding accessories such as air nozzles and handles)