服务热线:
热门关键词: 洁净检测仪器 净化设备 在线监测系统 过滤与分离
产品分类
PRODUCT CATEGORIES
全国服务热线
CLJ-3108T eight-channel laser dust particle counter
关键词:
所属分类:
咨询热线:
CLJ-3108T eight-channel laser dust particle counter
CLJ-3108T eight-channel laser dust particle counter
CLJ-3108T High-Flow Laser Dust Particle Counter
Product Overview
Light source: semiconductor laser
Sampling flow rate: 28.3L/min
Particle size channels: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0 μm or 0.3, 0.5, 0.7, 1.0, 5.0, 10.0, 15.0, 25.0 μm
Monitor: Large-screen LCD touch screen
Power supply: lithium battery, rechargeable 16.8V
Data storage: Capable of storing 20,000 sets of data
Communication: RS232, 9600 baud rate, USB interface
Printing function: thermal printer
Product Features
This instrument is a successful product jointly developed with Nanjing University of Science and Technology and other units in accordance with the requirements of the new version of GMP (2010 edition). The overall product is made of 316L stainless steel casing. The laser sensor utilizes a full semiconductor laser and a semiconductor photodiode receiver to ensure the stability of the light source and the accuracy of the signal receiver. This greatly reduces the stray scattering light within the scattering chamber, enhancing the signal-to-noise ratio of the sensor. It adopts an LCD touch screen for control, providing clear display of data and convenient operation. The instrument can automatically determine various purification levels based on the standards selected by customers.
Scope of application
This product is widely used in medical, pharmaceutical, electronics, chemical, optical, food, cosmetics, biological products, aerospace, and various testing institutions for the detection of dust particles
Main technical parameters:
Light source: Semiconductor laser, with a lifespan of over 10 years (more than 30,000 hours)
Sampling flow rate: 28.3L/min (1ft3) ±5%. It can automatically adjust the flow rate based on the length of the sampling tube, ensuring a constant flow rate despite resistance
Display mode: 155×85mm high-resolution LCD touch screen display with Chinese interface, real-time display, previous cycle display, real-time concentration display, display of time, date, measured value, temperature and humidity, room number, sampling point, sampling frequency, status, and other parameters. 95% UCL calculation, capable of directly displaying particle concentration (particles/cubic meter), with optional three-level permission management and audit tracking function.
Power Supply/Operating Power Supply: Maximum power consumption: 20W (1) Built-in long-lasting lithium battery: capable of continuous measurement for 3.5 hours, and an external charger can quickly charge it (<4 hours). The charger operates with a power supply of 100V~220V, 50/60Hz
Working hours: 8 hours
Counting mode: cumulative value, difference value, concentration value
Test methods: single, repeated, continuous, computational, remote
Unit conversion: The unit can be converted to m3
Particle size channels: 0.3, 0.5, 0.7, 1.0, 3.0, 5.0, 7.0, 10.0 μm or 0.3, 0.5, 0.7, 1.0, 5.0, 10.0, 15.0, 25.0 (μm) with simultaneous counting of eight particle size ranges. Particle size can be customized according to customer needs
Sampling period: 1~9999 (S) Delay count: 0~99 (S) Self-cleaning time: ≤10 (min)
Operating Environment: Temperature: 0~40℃ (50~104℉) Relative Humidity: 20~90%RH, no condensation Atmospheric Pressure: 86-106KPa
Temperature and humidity: optional (1) Temperature: 0~50℃±1℃. (2) Humidity: 0~100%RH±5%
UCL settings: Number of sampling points (A): Settable from 2 to 9 points. Sampling times per point (L): Settable from 2 to 9 times. Measurement position: 0-999
UCL report conforms to ISO14644-1, GB50073-2001, GMP (2010), and FS209E
Data storage: Capable of storing 20,000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, and data location port), with data not lost after power failure.
Communication interface: RS232, 9600 baud rate, USB interface
Alarm Settings: The instrument features a level alarm function, capable of sounding alarms for clean room levels exceeding the standards of 100, 1000, 10000, 300000, and 1000000. Additionally, it supports switching between four measurement standards, namely A/B/C/D, which align with the static and dynamic criteria of the latest GMP standards
Zero count Meets JJF-1190-2008 Dust Particle Counter Metrological Calibration Specification and ISO/FDIS21501-4:2018, with less than 1 particle every five minutes
Overlap error: 5%, at 2,000,000 grains per cubic foot
Printing function: Thermal printer
Weight: 8kg
Calibration: It can meet the requirements of JJF-1190-2008 "Calibration Specification for Dust Particle Counter", GB/T6167-2007 "Performance Test Method for Dust Particle Counter", and ISO/FDIS21501-4:2018(E) "Calibration Method for Light Scattering Dust Particle Counter in Clean Room". It can be traced back to the Shanghai Institute of Measurement and Testing Technology, and can also be calibrated by itself or by a third-party national metrology institution, with traceability to the National Institute of Standards and Technology (NIST). Our company has passed the national metrology standard establishment assessment, and our standard dust particle counter has also participated in the national dust particle counter value comparison conducted by metrology institutions in various provinces. It is used as a standard precision particle counter for relevant metrology and testing institutions during dust particle counter calibration work.
Standard configuration: professional isokinetic sampling head, sampling bracket with sampling tube, built-in printer, aluminum alloy case, operation manual, factory inspection report, zero counting filter, charger, sampling rack, printing paper
Overall dimensions: 220×220×230 (mm)
上一页