Y09-3016 Handheld laser dust particle counter
关键词:
所属分类:
咨询热线:
Y09-3016 Handheld laser dust particle counter
Y09-3016 Handheld laser dust particle counter
This instrument is the latest handheld laser dust particle counter developed by the company, which uses a full half body laser sensor. It can be connected to a PC computer data acquisition system for remote control and can directly observe the testing situation of the instrument. The test data can be analyzed and processed through a computer and saved as an Excel file. All technical indicators meet the requirements of the JJF1190-2008 calibration regulations issued by the State Administration of Metrology. The entire machine adopts American microcomputer control processing technology, semiconductor laser sensor technology, and imported air pump technology, which has the characteristics of multiple functions, high measurement accuracy, fast speed, easy portability, and simple operation. The instrument can simultaneously measure the number of dust particles of multiple sizes in one sampling, and this product has been widely used in clean room testing of electronic production enterprises; On site inspection and leak detection of filters; It can monitor biosafety, HVAC systems, computer rooms, beverage packaging environments, drug and medical device production environments, hospital clean operating rooms, automotive spraying environments, microelectronics, biochemical products, food hygiene, fine chemicals, precision machinery, aerospace and other production and scientific research departments. It is the preferred instrument for HVAC and pharmaceutical enterprises and their supervision and management departments to implement GMP standards and electronic production enterprises.
Main technical parameters:
Light source semiconductor laser, with a lifespan of over 10 years>30000 hours
Sampling flow rate 2.83L/min (0.1ft3)
Counting efficiency 0.3 μ m 50% ± 20 0.5 μ m 100% ± 10
Particle size concentration indication error ± 20%
Particle size distribution error ± 30%
Maximum sampling concentration 30000 particles/L
Display mode: 320 × 240 high-resolution LCD screen with Chinese and English interface selection, real-time display, previous cycle display, real-time concentration display, can display time, date, measurement value, temperature and humidity, room number, sampling point, sampling frequency, battery power, status and other parameters, 95% UCL calculation, can directly display particle concentration (particles/cubic meter)
Rechargeable lithium-ion battery, 16.8 V, 2600mAh
Power supply AC adapter, AC: 100V~245V, 50/60 Hz to DC: 16.8V, 1A;
Working time 8 hours (remaining battery indicator)
Count mode cumulative value, difference value, concentration value
The testing method is single, repetitive, continuous, computational, and remote
Unit conversion: Units can be converted to m3, ft3
The particle size channels are 0.3, 0.5, 1.0, 3.0, 5.0, and 10.0 μ m, and six particle sizes are counted simultaneously Optional particle sizes: 0.3 μ m, 0.5 μ m, 0.7 μ m, 1.0 μ m, 3.0 μ m, 5.0 μ m
Sampling period 1~9999 (S) Delay count: 0~99 (S) Self purification time: ≤ 10 (min)
Working environment temperature: 0-40 ℃ (50-104 ℉) Relative temperature: 20-90% RH, no condensation Atmospheric pressure: 86-106KPa
Temperature and humidity: Choose (1) Temperature: 0-50 ℃± 1 ℃ (2) Humidity: 0-100% RH ± 5%
UCL sets the number of sampling points (A): 2-9 points are set. Sampling frequency per point: (L) 2-9 times Set measurement position: 0-999
UCL report ISO14644-1, GB50073-2001, new version 2010 GMP, FS-209E
The data storage can store 1000 sets of data (circular buffer) (including particle size, data, environmental data, year, month, day, time, sampling volume, data location port), and the data will not be lost after power failure.
Communication interface RS232 9600 baud rate
The alarm setting instrument has a level alarm function, which can alarm for cleanroom levels 100, 1000, 10000, 100000, 300000, and 1000000 when the battery voltage exceeds the standard
Zero counting complies with JJF-1190-2008 Dust Particle Counter Measurement Calibration Specification and JIS B9921:1997, with less than one count every five minutes
Overlap error of 5%, at 2000000 particles per cubic foot
Printing function external printer (optional)
Dimensions 23X13X4.5 (cm)
Weight 0.6kg
The calibration specification for JJF-1190-2008 dust particle counter can be traced back to the National Institute of Standards and Technology (NIST) in the United States. Our company has passed the national metrological standard building assessment and can be traced back to the Shanghai Institute of Metrology and Measurement Technology, or can be calibrated by ourselves or by third-party national metrology institutions